86 Chapter: 6
4 For a Square Wave bolus do these steps:
The SET SQUARE BOLUS screen will appear.
Change the amount if desired. Press ACT to
accept.
For a Dual Wave bolus do these steps:
NOTE -
The number of units you program for the
SET DUAL BOLUS TOTAL is a total of both the
Normal and Square Wave bolus units.
a. The SET DUAL BOLUS TOTAL
screen will appear. Change
the amount if desired. Press
ACT to accept.
b. In the next screen, notice the
screen shows the Normal
(Now) and Square portions of
the Dual Wave bolus. Press
ACT to accept the portions
suggested by the Bolus Wizard
feature, or press
/
to
change these portions then
press ACT.
NOTE -
The Bolus Wizard
feature recommends splitting
the food portion of your bolus
50/50 between the Square
and Now portions. The entire
correction amount is always
recommended to the Now portion. In this example,
the Now portion consists of half of the food
insulin plus the correction amount less that active
insulin (1.5U+2.5U-1.5U). This gives 2.5U or 62%
of the total insulin of 4.0U. The square portion
consists of the other half of the food insulin
(1.5U) which is 38% of the total insulin of 4.0U.
SET SQUARE BOLUS
Estimate
ACT
4.0U
ACT
SET DUAL BOLUS TOTAL
Estimate
4.0U
ACT
2.5U
1.5U
Now:
62%
Square:
38%