What is a CHECK SETTINGS alarm?
This alarm occurs after an E alarm or after you clear your pump. It is advising you to make sure that all
your settings are correct. A CHECK SETTINGS alarm occurs after any of these actions:
• All user settings were cleared (set back to their defaults) because there was an E-error alarm,
• The CLEAR SETTINGS function was performed,
• Or after you rewind when practicing without a reservoir when your first get your pump. In this case, it
is just a reminder to make sure all your settings are programmed before you begin using the pump
with insulin.
My screen appears distorted
The screen may appear distorted or have a rainbow-like appearance if you are wearing polarized
sunglasses, are in bright sunlight, or in extreme high or low temperatures. If your screen appears
distorted:
• Take off your sunglasses.
• Move into the shade.
• Make sure your pump is not in direct heat (next to a heater) or cold (worn on the outside of your
clothing on a very cold day).
• Do not return the pump: this is a normal property of this type of screen on any device.
I cannot get out of the priming loop
There are several causes of and resolutions to this problem:
1 Is the reservoir pump full?
• If no, place a full reservoir or shipping cap in the pump.
• If yes, make sure you are disconnected from the pump.
2 Hold the ACT button until the second set of beeps and the numbers appear on the screen.
• If yes, your pump is okay. Go to step 4 in the Manual prime section in the Starting on insulin
chapter to finish the manual prime.
• If you did not hear a second set of beeps or numbers did not appear on the screen, change your
infusion set and repeat this step.
3 If you still do not hear the beeps and see the number count on the screen, call the helpline or your
local representative.
Troubleshooting and alarms 131
Chapter 11