balloon may lead to inaccurate and
potentially dangerous gas readings.
Use of non-standard
calibration gas and/or calibration kit
components when calibrating the
Cannonball 3 may lead to dangerously
inaccurate readings and may void the
standard Sperian Instrumentation’s
warranty.
3.2.1 Reading “Too High” or “Too
Low” for zero adjust
To reduce the chances of the Cannonball
3 being inadvertently calibrated in
contaminated air, only small adjustments
are allowed in the automatic fresh air
calibration sequence. If the necessary
adjustments are too large, the display will
indicate the sensor (or sensors) affected,
and a message screen will indicate that
the reading is “Too Low” or “Too High”
for zero adjustment. In this case the
instrument must be fresh air calibrated
using the manual calibration procedures
as discussed in section 3.2.1 of this
manual.
Once the instrument has been
successfully fresh air calibrated using the
manual fresh air calibration procedure,
subsequent calibration adjustments may
be made using the MODE button and
automatic calibration logic discussed in
this section.
3.2.2 Automatic span calibration
with more than one gas cylinder
Depending on the sensor configuration of
the Cannonball 3, it will sometimes be
necessary to use more than one cylinder
of calibration gas to complete the
calibration of the Cannonball 3.
Note: When using the calibration
balloon and multiple cylinders of
calibration gas during calibration, it is
necessary to purge the calibration
balloon and refill it with new
calibration gas between span
adjustments. In this case the display
will indicate the type and
concentration of the next cylinder of
calibration gas to be applied.
If the initial span calibration covered
above fails to calibrate the LEL sensor or
any of the toxic sensors, the Cannonball
3 will show the following screen, which
indicates the calibrations that still need to
be performed.
Apply one of these gases
25.0 ppm H
2
S
10.0 ppm SO
2
Once the instrument detects the proper
gas and the reading stabilize, the
instrument will automatically calibrate the
sensor.
SO
2
Sensor Adjusted to 10.0ppm
(SO
2
Max Possible to 37)
When all sensors currently recognized
have been successfully span-calibrated,
the instrument will display “Auto
Calibration Completed” and proceed to
shut itself off.
Remove all fittings from the Cannonball 3
before turning the instrument back on.
Note: It is possible to exit the
automatic span calibration sequence
at any time prior to completion by
pressing and holding down the MODE
button to turn the instrument off. The
instrument will retain the updated
settings for those sensors whose
span adjustments have been
successfully completed. Sensors that
were not successfully adjusted at the
time the auto-calibration sequence
was terminated will trigger a “Needs
Cal” message at the time the
instrument is next turned on. The
accuracy of those remaining sensors
should be verified by exposure to
known concentration test gas before
the instrument is put back into
service.