44 Chapter 4
Normal bolus during a square wave bolus
Use this feature if you need additional insulin. You can program a
normal bolus during the delivery of a square wave bolus. The nor-
mal bolus will interrupt the square wave bolus that is being deliv-
ered. The square wave delivery will continue once the normal bolus
delivery is finished.
1. From the HOME screen, press ACT. Suspend will be selected.
Select Bolus. Press ACT.
2. Set Bolus will be selected. Press ACT.
3. The screen will now show that a square wave bolus is in
progress. Press ACT.
MiniMed
MAIN MENU
Suspend
Basal
Bolus
MiniMed
BOLUS MENU
Bolus History
Max Bolus
Set Bolus
MiniMed
SQUARE: Set
Delivered
Time Left
0.2U
1:29
ACT to Program
0.6U